banner

News

Aug 30, 2023

Feasibility of in vivo swine models using guide wire

Scientific Reports volume 13, Article number: 7185 (2023) Cite this article

285 Accesses

Metrics details

Several in vivo swine models of benign biliary stenosis (BBS) have been recently reported for preclinical studies of novel endoscopic techniques and devices. The aim of this study was to evaluate the efficacy and feasibility of large animal models of BBS by using intraductal radiofrequency ablation (RFA) assisted by guide wire. Six in vivo swine models were made by using an intraductal RFA for cauterization at 10 W, 80 °C, 90 s in the common bile duct (CBD). Endoscopic retrograde cholangiopancreatography (ERCP) was performed with cholangiography and histologic evaluation was done for the common bile duct. Blood tests were examined before, after, and at the final follow-up. Guide wire assisted RFA electrode produced BBS in all (6/6, 100%) animal models without severe complications. Fluoroscopy findings at 2 weeks after intraductal RFA in every model revealed BBS in the common bile duct. In histologic evaluations, fibrosis and chronic inflammatory changes were noted. After the procedure, ALP, GGT, and CRP were elevated and decreased after an appropriate drain. A swine model of BBS is developed by inducing intraductal thermal injury using intraductal RFA assisted by guide wire. This novel technique for inducing BBS in swine is effective and feasible.

Endoscopic biliary drainage has been used as an initial treatment for benign biliary stricture (BBS). Recently, multiple plastic stents and covered self-expandable metal stents have been reported to show promising outcomes for the management of BBS1,2. However, adequate biliary drainage over the long-term using biliary stents is still challenging. Many biliary endoscopic devices including biliary stents have been recently developed to improve the patency of devices3,4,5. However, these attempts to increase the long-term drainage effect by changing the shape or material of plastic stents do not have satisfactory effects. Therefore, there are increasing requirements for new devices that can improve limitations of conducting endoscopic management in BBS.

Before applying to endoscopic devices in humans, appropriate animal models are important for preclinical study to evaluate the efficacy of endoscopic devices. In the past, previous studies mostly reported surgically created percutaneous approaches6,7,8. However, procedures of making animal models using surgically created percutaneous approaches were laborious. Recently, several studies have reported the development of in vivo and in vitro animal models for BBS using an endoscopic biliary approach9,10,11. The first endoscopic biliary approach by Rumella et al. used heat probe and multipolar probe to produce BBS12. However, these studies were conducted using methods including heat probe and multipolar probe other than intraductal radiofrequency ablation (RFA) for BBS. Even if an endoscopic approach was used, the experiment was conducted using a non-thermal injury method such as endoscopic detachable snare. Some studies have reported a short observation time without performing follow-up liver function tests commonly conducted in clinical practice.

Comparing other thermal injury methods for animal models of BBS, using intraductal RFA device is comfortable and easy to control energy dose (W), temperature (°C), and exposure time (sec). Using guide wire methods, intraductal RFA devices might more easily control the location and degree of bile duct injury. However, there were only a few investigations for intraductal application of RFA. In addition, an effective and safe energy dose for intraductal thermal injury with RFA is still not established. Thus, the aim of this study to assess the development of reproducible large animal models of BBS using endo biliary RFA and to investigate an effective and safe energy option for application in producing BBS.

In all six swine animal models, we succeeded in generating BSS using intraductal RFA without any complications such as bleeding or perforation (success rate = 100% (6/6), severe complication rate = 0%). Blood levels of WBC, AST, ALT, ALP, GGT, and CRP of all experimental animals were measured before the intraductal RFA procedure, after the intraductal RFA procedure (2 weeks after RFA), and before euthanizing animals (Figs. 1, 2, 3A–G). Blood levels of WBC, AST, ALT, ALP, GGT, and CRP were elevated after the intraductal RFA procedure but decreased after biliary stenting. Figure 1H,I are biliary fluoroscopy findings at 2 weeks after RFA in experimental animals 1 and 2, demonstrating biliary stenosis. The same trend of blood test results as in Fig. 1 was observed in Fig. 2. The difference between Figs. 1 and 2 was that the follow-up was performed at 3 months after biliary stenting in Fig. 2. Biliary stenosis was confirmed by fluoroscopy findings at 2 weeks after RFA in experimental animals 3 and 4 (Fig. 2H,I). Experimental animals 5 and 6 were followed for 5 months after biliary stenting. They showed the same tendency of blood test results (Fig. 3A–G). Biliary stenosis was also confirmed by fluoroscopy findings at 2 weeks after RFA in experimental animals 5 and 6 (Fig. 3H,I).

Blood test and cholangiography of biliary stricture pre-RFA, post-RFA, and at 1-month follow-up. (A) WBC changes; (B) T. bilirubin changes; (C) AST changes; (D) ALT changes; (E) ALP changes; (F) GGT changes; (G) CRP changes of 1 month follow-up pigs; (H) Cholangiography of biliary stricture at 1-month follow-up (pig 1); (I) Cholangiography of biliary stricture at 1-month follow-up (pig 2). WBC white blood cell count, T. bilirubin total bilirubin, AST aspartate transaminase, ALT alanine transaminase, ALP alkaline phosphatase, GGT gamma-glutamyl transferase, CRP C-reactive protein.

Blood test and cholangiography of biliary stricture pre-RFA, post-RFA, and at 3-month follow-up. (A) WBC changes; (B) T. bilirubin changes; (C) AST changes; (D) ALT changes; (E) ALP changes; (F) GGT changes; (G) CRP changes of 3-month follow-up pigs; (H) Cholangiography of biliary stricture of 3-month follow-up (pig 3); (I) Cholangiography of biliary stricture of 3-month follow-up (pig 4). WBC white blood cell count, T. bilirubin total bilirubin, AST aspartate transaminase, ALT alanine transaminase, ALP alkaline phosphatase, GGT gamma-glutamyl transferase, CRP C-reactive protein.

Blood test and cholangiography of biliary stricture pre-RFA, post-RFA, and at 5-month follow-up. (A) WBC changes; (B) T. bilirubin changes; (C) AST changes; (D) ALT changes; (E) ALP changes; (F) GGT changes; (G) CRP changes of 5-month follow-up pigs; (H) Cholangiography of biliary stricture of 5-month follow-up (pig 5); (I) Cholangiography of biliary stricture of 5-month follow-up (pig 6). WBC white blood cell count, T. bilirubin total bilirubin, AST aspartate transaminase, ALT alanine transaminase, ALP alkaline phosphatase, GGT gamma-glutamyl transferase, CRP C-reactive protein.

The common bile duct (CBD) diameter was 2.5 ± 0.5 mm in macroscopic findings. The biliary stricture length measured after harvesting of experimental animals was 36 mm ± 0.5 mm in macroscopic findings. Using H&E-stained tissue sections, the degree of histological damage was compared by examining the depth of inflammation, the degree of deposition of neutrophils, the presence of mucosal ulceration, and the overall score (Figs. 4, 5 and Tables 1, 2). The total histological score indicating histological damage of plastic stents was observed to be higher for the 3-month point than that for the 1-month point total score, median (range) of 1-month pigs vs. 3-month pigs: 6 (6–8) vs. 6.5 (6–8), although their difference was not statistically significant (p = 0.057) (Fig. 4 and Table 1). However, in the animal group in which the plastic stents were mounted for 5 months, the total score was observed to be significantly higher than those mounted for less time (total score, median (range) of 3-month pigs vs. 5-month pigs: 6.5 (6–8) vs. 9 (9–9), p = 0.029) (Fig. 5 and Table 2). The degree of tissue damage was compared using immunofluorescent staining (Fig. 6). From the top to the bottom, photomicrographs of H&E, Masson Trichrome staining, and immunofluorescence staining using CK 19 are shown. H&E staining revealed that the deposition degree of neutrophils of the tissue in the contact area of the plastic stents (Fig. 6). When the same area was observed after Masson Trichrome staining, extensive fibrosis of the tissue in the contact area was observed. At the bottom of Fig. 6, the immunofluorescent staining was performed using CK 19 for histologic evaluation.

Comparison of microscopic images of bile duct inflammation using histologic scoring system between 1-month (left) and 3-month (right) follow-up animal models.

Comparison of microscopic images of bile duct inflammation using histologic scoring system between 3-month (left) and 5-month (right) follow-up animal models.

Microscopic images of plastic stented bile ducts from the swine animal model. H–E, M-T staining and CK 19 immunohistochemistry (from the top); H–E hematoxylin eosin, M-T Masson trichrome, CK 19 cytokeratin 19.

The aim of this study was to develop an in vivo swine model of BBS using guide wire assisted intraductal RFA. Animal experiments using in vivo bile ducts are important for preclinical tests during the development and improvement processes of pancreaticobiliary plastic stents technology. According to the literature, ERCP or laparoscopic surgery can be performed to reach the duodenal papilla and form a biliary stenosis model using detachable snare11. To form a biliary stenosis model, an intraductal heat probe through laparotomy or an intraductal radiofrequency heat therapy electrode can be used10. In this study, biliary stenosis models were successfully prepared using radiofrequency thermal therapy electrodes in the biliary tract through ERCP without post-procedure complications such as cholangitis or post ERCP pancreatitis. It is expected to be of great help in future preclinical research.

In this study, we performed H&E staining, Masson Trichrome staining, and immunofluorescence staining (CK 19) to achieve an objective histopathological evaluation. Furthermore, the veterinarian pathologist in the animal laboratory evaluated and calculated the histological score based on the histological scoring system shown in Supplementary Table 1 commonly used in previous studies13,14,15. Histological damage caused by plastic stents was found to be higher when the follow-up time was longer. Interestingly, the total histologic score did not show any significant differences between different time points less than 3 months after inserting biliary plastic stents. However, at 5 months after stent insertion, a statistically significant difference in histological score was observed compared to that at shorter follow-up. Considering our results and cost effectiveness of the animal study, 1-month point after RFA seems to be the best timing for testing the new devices in the BBS in vivo swine models. Therefore, in clinical practice, it is recommended to remove the plastic stent between 3 and 6 months, which is the average patency period in previous studies16,17,18 to avoid histological damage.

Our study has several limitations. First, this was a preclinical study using animal experiments. Results of the experiment cannot be generalized to human responses of patients with benign biliary stenosis disease. Second, the anatomy of bile ducts in swine is different from that in human. In swine, unlike human, the pancreatic duct and bile duct are separated. Finally, the efficacy and the feasibility were evaluated using a small number of experimental animals. Nevertheless, this study has its merit. It introduces the development of an experimental animal model of biliary stenosis with a reproducible, safe, and reliable method using radiofrequency heat therapy electrodes in the biliary tract.

In conclusion, the efficacy and safety of in vivo swine models of BBS using guide wire assisted intraductal RFA were confirmed through a preclinical laboratory animal study. In the future, a good method that can be used to evaluate the efficacy and safety of biliary plastic stents in the human body needs to be developed.

A total of six female micro pigs (Micro pig M-type; Medi Kinetics Co., Ltd, Pyeongtaek, Gyeonggi-do, Korea) with an average weight of 50 kg at 10–12 weeks old were selected for animal experiments. Before starting the experiment, they were acclimated for 1 week. Only healthy animals were used for these experiments. All were bred in an animal breeding room set at a temperature of 23 ± 2 °C, a relative humidity of 50 ± 5%, a ventilation frequency of 10–12 times/hour, an illumination time of 08:00 to 20:00, and an illumination intensity of about 400 lx. During the quarantine period and experimental period, each pig was placed in one cage. Solid feed (Purina) was supplied twice a day, once before the start of day and once at 4:00 pm. Experiments were conducted after fasting for 24 h the day before the procedure. This study was carried out in accordance with relevant guidelines and regulation of the Animal Experimental Ethics Committee of the Samsung Life Science Research Institute, an accreditation body for AAALAC International (Association for Assessment and Accreditation of Laboratory Animal Care International) (IACUC Approval Number: 20160712001) and the Animal Research: Reporting of In Vivo Experiments guidelines (ARRIVE guidelines (https://arriveguidelines.org)) were followed for the in-vivo studies.

A total of six female micro pigs were randomly assigned to three groups (2 pigs per group) for 1 month, 3-month, and 5-month monitoring. The biliary stenosis model using electrode biliary cauterization was performed based on a previously published method10. Briefly, after fasting for 24 h, pigs were sedated by a veterinarian using ketamine® 50 mg/ml 20 mg/kg, intramuscular injection of zolazepam (Zoletil®; 6 mg/kg), and xylazine (Rompun®; 2 mg/kg). Tracheal intubation was then performed. After tracheal intubation, anesthesia was maintained using 2% isoflurane. Electrocardiogram, heart rate, blood pressure, oxygen saturation, and end-tidal CO2 were monitored by the veterinarian. Enrofloxacin (2.5 mg/kg) was intramuscularly administered before the procedure and at 2 days after the procedure to prevent cholangitis due to the procedure. On the day of the procedure, ketoprofen (2 mg/kg) was intramuscularly administered for pain control. Endoscopic retrograde cholangiopancreatography (ERCP) was conducted using TJF240 (Olympus America, Inc, Melville, NY, USA), a therapeutic endoscope. First, we found and observed the duodenal papilla (Supplementary Fig. 1A). The ERCP was performed using the wire-guided cannulation method, which could selectively cannulate the bile duct using a guide wire under a fluoroscope (Supplementary Fig. 1B). After that, the Ampullae of Vater (AOV) was expanded with a Hurricane balloon catheter (Boston Scientific Corp., 10 mm in diameter) along the guide wire. The intra-biliary radiofrequency ablation (RFA) electrode was then inserted into the biliary tract along with the guide wire (Supplementary Fig. 1C,D). The RFA electrode (ELRA electrode, 7–10-W/33-mm type, STARmed Co. Ltd, Goyang, Gyeogi-do, Korea) mounted in the CBD was used for cauterization at 10 W and 80 °C for 90 s19,20,21. For maximizing the contact surface between the CBD and RFA electrode, the suction technique was used21.

After 2 weeks of intraductal RFA, ERCP was performed to confirm biliary stenosis of treated experimental animals. Blood tests including White Blood Cell (WBC), Aspartate Transaminase (AST), Alanine Transaminase (ALT), Alkaline Phosphatase (ALP), Gamma-Glutamyl Transferase (GGT) and C-Reactive Protein (CRP) were performed before the procedure, after the procedure, and at the final follow-up. Using a 0.035-inch guidewire (Hydrophilic Tipped Guidewire, Boston Scientific Corp., Natick, MA, USA) under fluoroscopy, two commercially available 10-F biliary plastic stents (polyurethane tube of outer diameter 3.3 mm and inner diameter 2.0 mm; polyethylene tubing of outer diameter 3.3 mm and inner diameter 2.0 mm) were placed in the biliary tract (Fig. 7). The proximal end of the plastic stent was mounted to be located in the intrahepatic bile ducts of different branches (Fig. 7C).

Fluoroscopy in swine models of biliary stenosis and plastic stent insertion. (A) Swine model of biliary stenosis (arrow) was confirmed by cholangiography images. Stenosis of the common bile duct with upstream dilatation; (B) Plastic stents were successfully inserted into swine bile duct; (C) Plastic stents were inserted into the dilated bile duct of swine. Proximal end of the plastic stent was placed in a different branch of the intra hepatic bile duct (arrows).

At 1 month, 3 months, and 5 months after plastic stents were inserted, 2 pigs at each time point were injected intramuscularly ketamine (Ketamine®; 50 mg/ml 20 mg/kg), zolazepam (Zoletil®; 6 mg/kg), and xylazine (Rompun®; 2 mg/kg) by a veterinarian on the same day of the procedure for sedation. After sedation, tracheal intubation was performed. Anesthesia was maintained using 2% isoflurane. Electrocardiogram, heart rate, blood pressure, oxygen saturation, and end-tidal CO2 were monitored by the veterinarian. Open laparotomy of all pigs was performed by one very skilled veterinarian. A median incision was made, and the duodenum was removed.

Pigs were euthanized at 1 month, 3 months, and 5 months after biliary plastic stents were inserted. Histopathological examination was then performed. After the liver, biliary tract, gallbladder, and duodenum were removed from experimental animals, a large amount of KCl (Potassium Chloride) was injected to induce euthanasia. From the extracted liver, biliary tract, gallbladder, duodenum, and AoV were dissected to the proximal end of the plastic stent located in the right and left intrahepatic bile duct. After the dissection, an incision was made in the longitudinal direction to confirm the proximal intrahepatic bile duct and distal biliary duct stenosis. The bile duct tissue was incised from the intrahepatic bile duct into which different plastic stents had been inserted and cut into sections, followed by H&E staining and Masson Trichrome staining. Histopathological examination was performed to reinforce the histological scoring method of previous studies. Scoring was performed in a blinded method using H&E-stained tissue sections. All sections were classified according to the extent of inflammation (1) (1, minimal; 2, confined to mucosa; and 3, extended to muscle with neutrophil infiltration). The severity of neutrophil infiltration was scored (2) as 0 for Absent, 1 for Mild, 2 for Moderate, and 3 for Severe. The presence of mucosal ulceration (3) was scored as 0 for Absent, 2 for Local, and 4 for Diffuse (Supplementary Table 1)13,14,15. Based on the histological scoring system shown in Supplementary Table 1, the histological score was calculated by a veterinarian pathologist in the animal laboratory of Samsung Life Sciences Research Institute.

All experiments were performed or measured three or more times to compare the histopathology score. IBM SPSS version 24.0 (IBM Corp., Armonk, NY, USA) was used for all statistical analyses.

The datasets of this study are available. All result generated or analyzed by the raw dataset are included in this published article and Supplementary Information files. Pus, the raw datasets generated during the study are available for research purposes from the corresponding author on reasonable request.

Benign biliary stenosis

Radiofrequency ablation

Endoscopic retrograde cholangiopancreatography

White blood cell

Aspartate transaminase

Alanine transaminase

Alkaline phosphatase

Gamma-glutamyl transferase

C-reactive protein

Hematoxylin eosin

Masson trichrome

Cytokeratin 19

Gupta, R. & Reddy, D. N. Stent selection for both biliary and pancreatic strictures caused by chronic pancreatitis: Multiple plastic stents or metallic stents?. J. Hepatobiliary Pancreat. Sci. 18, 636–639 (2011).

Article PubMed Google Scholar

Poley, J.-W., van Tilburg, A. J., Kuipers, E. J. & Bruno, M. J. Breaking the barrier: Using extractable fully covered metal stents to treat benign biliary hilar strictures. Gastrointest. Endosc. 74, 916–920 (2011).

Article PubMed Google Scholar

Terruzzi, V. et al. Prospective randomized trial comparing Tannenbaum Teflon and standard polyethylene stents in distal malignant biliary stenosis. Gastrointest. Endosc. 51, 23–27 (2000).

Article CAS PubMed Google Scholar

England, R. et al. A prospective randomised multicentre trial comparing 10 Fr Teflon Tannenbaum stents with 10 Fr polyethylene Cotton-Leung stents in patients with malignant common duct strictures. Gut 46, 395–400 (2000).

Article CAS PubMed PubMed Central Google Scholar

Chan, G. et al. The role of ciprofloxacin in prolonging polyethylene biliary stent patency: A multicenter, double-blinded effectiveness study. J. Gastrointest. Surg. 9, 481–488 (2005).

Article PubMed Google Scholar

Vorwerk, D., Guenther, R. W., Kuepper, W. & Kissinger, G. Thermal laser-induced stenosis of the common bile duct an alternative model for experimental research. Investig. Radiol. 24, 758–761 (1989).

Article CAS Google Scholar

Posner, M. C. et al. A model of reversible obstructive jaundice in the rat. J. Surg. Res. 48, 204–210 (1990).

Article CAS PubMed Google Scholar

Hirazawa, K. et al. New technique for inducing reversible obstructive jaundice in the rat. Eur. Surg. Res. 29, 195–201 (1997).

Article CAS PubMed Google Scholar

Park, J.-S., Jeong, S., Kim, J. M., Park, S. S. & Lee, D. H. Development of a swine benign biliary stricture model using endoscopic biliary radiofrequency ablation. J. Korean Med. Sci. 31, 1438–1444 (2016).

Article CAS PubMed PubMed Central Google Scholar

Shin, J. U. et al. Intraductal thermal injury using a heat probe and radiofrequency ablation electrode in a swine model of biliary stenosis. Clin. Res. Hepatol. Gastroenterol. 37, 159–165 (2013).

Article PubMed Google Scholar

Kwon, C.-I. et al. Bile flow phantom model and animal bile duct dilation model for evaluating biliary plastic stents with advanced hydrophilic coating. Gut Liver 10, 632 (2016).

Article CAS PubMed PubMed Central Google Scholar

Rumalla, A. et al. Development of a swine model for benign stenosis of the bile duct by endoscopic application of intraluminal thermal injury. Gastrointest. Endosc. 57, 73–77 (2003).

Article PubMed Google Scholar

Woo, Y. S. et al. Novel flower-type covered metal stent to prevent cholecystitis: Experimental study in a pig model. Surg. Endosc. 30, 1141–1145 (2016).

Article PubMed Google Scholar

Lee, S. S. et al. Histological changes in the bile duct after long-term placement of a fully covered self-expandable metal stent within a common bile duct: A canine study. Clin. Endosc. 47, 84–93 (2014).

Article CAS PubMed PubMed Central Google Scholar

Kjaer, D. W., Mortensen, F. V., Møller, J. K., Hamilton-Dutoit, S. J. & Funch-Jensen, P. Internal gallbladder drainage prevents development of acute cholecystitis in a pig model: A randomized study. Ann. Surg. Innov. Res. 4, 1–7 (2010).

Article Google Scholar

Gilbert, D. et al. Status evaluation: Biliary stents. American Society for Gastrointestinal Endoscopy Technology Assessment Committee. Gastrointest. Endosc. 38, 750–752 (1992).

CAS PubMed Google Scholar

Kadakia, S. C. & Starnes, E. Comparison of 10 French gauge stent with 11.5 French gauge stent in patients with biliary tract diseases. Gastrointest. Endosc. 38, 454–459 (1992).

Article CAS PubMed Google Scholar

Wagh, M. S. et al. Multicenter randomized trial of 10-French versus 11.5-French plastic stents for malignant biliary obstruction. Diagn. Ther. Endosc. 2013, 891915 (2013).

Article PubMed PubMed Central Google Scholar

Cho, J. H. et al. Safety and effectiveness of endobiliary radiofrequency ablation according to the different power and target temperature in a swine model. J. Gastroenterol. Hepatol. 32, 521–526 (2017).

Article PubMed Google Scholar

Oh, D., Ligresti, D. & Seo, D.-W. Novel swine biliary dilatation model with temperature-controlled endobiliary radiofrequency ablation: An effective tool for training in EUS-guided biliary drainage. Endosc. Ultrasound 9, 245 (2020).

Article PubMed PubMed Central Google Scholar

Kwon, C.-I. et al. Optimal reproduction of a porcine benign biliary stricture model using endobiliary radiofrequency ablation. Sci. Rep. 12, 1–11 (2022).

Article ADS Google Scholar

Download references

We thank BCM for their dedication during the development of animal models. This study was partially supported by Sungkyunkwan University Research Fund.

Department of Internal Medicine, Kangnam Sacred Heart Hospital, Hallym University College of Medicine, Seoul, South Korea

Jae Keun Park

Division of Gastroenterology, Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, 81 Irwon-ro, Gangnam-gu, Seoul, 06351, South Korea

Ju-Il Yang, Joo Kyung Park, Kwang Hyuck Lee, Jong Kyun Lee & Kyu Taek Lee

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

Conceptualization, Writing—Original Draft: J.K.P. (Jae Keun Park), Methodology: K.T.L., Formal analysis: J.I.Y., J.K.P. (Joo Kyung Park), Investigation and Resources: K.H.L., J.K.L., K.T.L., Supervision: K.T.L. Approval of final manuscript: all authors.

Correspondence to Kyu Taek Lee.

The authors declare no competing interests.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and Permissions

Park, J.K., Yang, JI., Park, J.K. et al. Feasibility of in vivo swine models using guide wire-assisted intraductal radiofrequency ablation for benign biliary stricture. Sci Rep 13, 7185 (2023). https://doi.org/10.1038/s41598-023-33867-9

Download citation

Received: 04 October 2022

Accepted: 20 April 2023

Published: 03 May 2023

DOI: https://doi.org/10.1038/s41598-023-33867-9

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative

By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.